Skip to content
PROGIENE, Protection & Hygiène
Actosed Forte Ready 5 L jerrycan, with a tube of test strips and a sachet of neutraliser (sodium bisulfate)
InstrumentsHigh-level disinfection

Actosed Forte Ready

High-level disinfectant based on 2% glutaraldehyde.

Ready-to-use solution for the final disinfection by immersion of pre-cleaned, reusable, heat-sensitive semi-critical and critical medical devices, invasive and non-invasive.

Packaging and references
1 format

Overview

Applications

  • Cleaned flexible and rigid endoscopes
  • Stainless steel surgical instruments, including minimally invasive surgery
  • Anaesthesia equipment and non-heat-resistant instruments
  • Manual use or in an endoscope washer

Key benefits

  • Ready to use, no activator
  • Wetting power without foaming
  • Solution reusable for up to 15 days, checked with test strips
  • Wide material compatibility

Technical characteristics

Composition

2% glutaraldehyde buffered to neutral pH, with corrosion inhibitors and auxiliary substances.

Material compatibility

Glass, ceramic, stainless steel, hard rubber and synthetic materials: hard plastic, rubber (including latex and silicone), polyacrylate, polycarbonate.

Efficacy and standards

  • Bactericidal15 min

    EN 14561

    Ready to use

  • Fungicidal15 min

    C. albicans, A. brasiliensis

    EN 14562

    Ready to use

  • Tuberculocidal, mycobactericidal15 min

    EN 14348EN 14563

    Ready to use

  • Virucidal15 min

    Adeno, Polio, MNV

    EN 14476

    Ready to use

  • Sporicidal60 min

    EN 13704

    Ready to use

Instructions for use

Dilution or dose
Ready to use
Contact time
According to the required activity (see efficacy table)
  1. Pre-clean surfaces and channels with a suitable enzymatic cleaner (the manufacturer recommends Actoclean Perfect), rinse and dry.
  2. Pour the solution into the bath and immerse the instruments so that all surfaces and channels are in contact; keep the bath closed.
  3. Observe the contact time for the required activity.
  4. Rinse in water of at least drinking quality, fully immersing the device for at least 1 minute; repeat twice more, then dry.
  5. The solution remains usable for up to 15 days: check the minimum effective concentration with test strips and replace it sooner if it becomes cloudy or soiled.
  6. In an endoscope washer, follow the machine manufacturer’s instructions.

Packaging and references

FormatReference
5 L jerrycan01.0303.5

Documentation

Safety and storage

Use products safely. Always read the label and the safety data sheet before use.

Storage

Keep the container tightly closed in a cool, well-ventilated place. Between 5 °C and 40 °C. Keep only in the original container, with containment for any leaks. Do not store with oxidising agents; keep away from food and drink.

A quotation, a datasheet or advice?